Features
Testing efficiency
- The series lists eight parameters: K⁺, Na⁺, Cl⁻, Ca²⁺, pH, n-Ca, T-Ca and Li⁺.
- Results in 25 seconds, with a throughput of 40 tests per hour.
Electrode performance
- Sensitive electrodes designed for a long service life.
- The manufacturer lists a one-year electrode warranty.
Reagent options and cost
- Choose between an all-in-one reagent cartridge and bottled reagents.
- Designed for low consumable cost per test.
- Automatic standby reduces reagent use.
Operation and connectivity
- Automatic flushing of the fluid path.
- Barcode reader support.
- Test data management and connectivity with a laboratory information system (LIS).
Product specifications
| Item | Specification |
|---|---|
| Result time and throughput | 25 seconds per result; 40 tests per hour |
| Sample volume | 60–150 µL |
| Test method | Ion-selective electrode (ISE) |
| Sample types | Serum, plasma, whole blood, cerebrospinal fluid and diluted urine |
| Dimensions and net weight | 420 × 320 × 280 mm; 7.5 kg |
| Power supply | 100–240 V AC, 50/60 Hz |
| Display | 5-inch IPS touchscreen |
| Operating conditions | 5–40 °C; ≤85% RH; 760–1,060 hPa |
| Communication | LIS support |
| Interface | RS232-C |
| Printer | Built-in thermal printer |
| Accessory | Barcode scanner |
Technical parameters
Measurement range and accuracy
| Parameter | Accuracy (B) | Precision (CV) | Linearity | Stability (R) | Carryover | ||
|---|---|---|---|---|---|---|---|
| Range | Deviation (D) | Correlation (r) | |||||
| K⁺ | ≤3.0% | ≤1.0% | 1.50–7.50 mmol/L | ≤3.0% | ≥0.995 | ≤2.0% | ≤1.5% |
| Na⁺ | ≤3.0% | ≤1.0% | 100.0–180.0 mmol/L | ≤3.0% | ≥0.995 | ≤2.0% | ≤1.5% |
| Cl⁻ | ≤3.0% | ≤1.0% | 80.0–160.0 mmol/L | ≤3.0% | ≥0.995 | ≤2.0% | ≤1.5% |
| Ca²⁺ | ≤5.0% | ≤1.0% | 0.50–2.50 mmol/L | ≤5.0% | ≥0.995 | ≤3.0% | ≤2.0% |
| Li⁺ | ≤5.0% | ≤1.0% | 0.50–2.50 mmol/L | ≤3.0% | ≥0.995 | ≤2.0% | ≤2.0% |
| pH | ≤3.0% | ≤3.0% | 7.00–9.00 (unitless) | ≤5.0% | ≥0.995 | ≤3.0% | ≤2.0% |
Model configurations
- DN-X3 K⁺, Na⁺, Cl⁻
- DN-X4 K⁺, Na⁺, Cl⁻, Li⁺
- DN-X5 K⁺, Na⁺, Cl⁻, Ca²⁺, pH
- DN-X6 K⁺, Na⁺, Cl⁻, Ca²⁺, pH, Li⁺
Sample types
Serum, plasma, whole blood, cerebrospinal fluid and diluted urine
Test procedure
- Power on the analyzer.
- Run a two-point calibration.
- Run daily quality control (QC).
- Prepare the sample.
- Start the test.